From: Wayne Reiersen [reiersen@pppl.gov] Sent: Wednesday, September 24, 2003 3:48 PM To: Brad Nelson; Dave Williamson Subject: FW: CTE test of conductor I think this looks consistent with the CTD data. Wayne T. Reiersen Princeton Plasma Physics Laboratory MS08 PO Box 451 Princeton NJ 08543 609/243-2479 (phone) 609/243-3030 (fax) -----Original Message----- From: Steve Raftopoulos [mailto:sraftopo@pppl.gov] Sent: Wednesday, September 24, 2003 3:08 PM To: jchrzano@pppl.gov Cc: wreiersen@pppl.gov; nelsonbe@ornl.gov; ekearns@pppl.gov Subject: CTE test of conductor Jim, We set up an experiment to determine the Coefficient of Thermal Expansion (CTE) of the copper/epoxy composite. We measured the contraction of an impregnated bar when cooled from room to liquid nitrogen temperature (294 to 77 K). The following data was recorded. Room temperature - 21C (294 K) Room temperature length of bar - 32.424" LN2 temperature length of bar - 32.323" The calculated CTE is 1.435 x10E-6 (in/in per degree Celsius) Steve Raftopoulos Project Engineer Princeton Plasma Physics Laboratory P.O. Box 451, Princeton, NJ 08543 (609) 243-3626 Voice (609) 243-3248 Fax